Detail of beamline is described in "PF-XAFS Website"
Techniques
XAFS(X-ray Absorption Fine Structure)
- Energy range: 4 - 23 keV
- Transmission mode: QXAFS
- Fluorescent mode: Lytle detector
Other experimental with monochromatized X-ray
- Please contact us before experiment
Light source
PF Bending magnet #9
- Ring energy: 2.5 GeV
- Max. current: 450 mA (Usually Top-up operation)
- Critical energy: 4 keV
- Horizontal acceptance: 3 mrad
Beamline optics
Monochromator
- Si(111) Water-cooled double crystal
- Energy range: 4 - 23 keV
- Cam-drive (Kohzu system)
Focusing system
- Bent cylinder mirror (Rh coated)
- Standard beam size: 0.8H x 0.6V mm
Focused and Monochromatized mode
- Energy range: 4 - 23 keV
- Energy resolution: ΔE/E~2×10-4 (9 keV)
- Photon flux at sample position: 1x1011 phs/s (8 keV)
Experimantal station
Experimental hutch
- Hatch size: 3.2L x 3.15W x 2.6H m
- Door size: 2.7W x 2.5H m
Standard setup
- Flat table size: 2000L x 900W mm
- XAFS setup: Ionization chumber, Cryocooler
- Gas flow and sample heating system for in-situ experiment
Beamline staff
Beamline manager
- 阿部 仁 / ABE Hitoshi (KEK-PF)
hitoshi.abe[@]kek.jp
Sub manager
- 仁谷 浩明 / NITANI Hiroaki (KEK-PF)
hiroaki.nitani[@]kek.jp
- 丹羽 尉博 NIWA Yasuhiro (KEK-PF)
yasuhiro.niwa[@]kek.jp
- 松岡 亜衣 / MATSUOKA Ai (KEK-PF)
ai.kamijo[@]kek.jp
Menu